The influence of patient position on withdrawal force of thoracic epidural catheters

  • Zeropain Pain Treatment Clinic
  • Seoul National University
  • Sungkyunkwan University

Research output: Contribution to journal › Article › peer-review

Abstract

Summary We investigated the forces required to remove thoracic epidural catheters to determine the effect of patient position on removal. Eighty-four patients undergoing open thoracotomy and thoracic patient-controlled epidural analgesia were enrolled. Catheterisation was performed under fluoroscopic guidance before surgery, and the patients were allocated to one of three position groups for removal: prone; sitting; and lateral. On the third postoperative day, the peak tension during withdrawal in the assigned position was measured. No differences in mean (SD) forces were found between groups: prone 1.61 (0.39) N, supine 1.62 (0.61) N and lateral 1.36 (0.56) N (p = 0.140). The withdrawal forces required to remove thoracic epidural catheters were not affected by the position. Thus, the position for removal can be determined by patient's choice and clinical judgement. Anaesthesia

Original languageEnglish
Pages (from-to)19-22
Number of pages4
JournalAnaesthesia
Volume67
Issue number1
DOIs
StatePublished - Jan 2012