Radiation exposure to operator and patients during cardiac electrophysiology study, radiofrequency catheter ablation and cardiac device implantation procedures

C. H. Lee, J. H. Cho, S. J. Park, J. S. Kim, Y. K. On, J. Huh

Research output: Contribution to journalArticlepeer-review

Abstract

The purpose of this study was to measure the radiation exposure to operator and patient during cardiac electrophysiology study, radiofrequency catheter ablation and cardiac device implantation procedures and to calculate the allowable number of cases per year. We carried out 9 electrophysiology studies, 40 radiofrequency catheter ablation and 11 cardiac device implantation procedures. To measure occupational radiation dose and dose-area product (DAP), 13 photoluminescence glass dosimeters were placed at eyes (inside and outside lead glass), thyroids (inside and outside thyroid collar), chest (inside and outside lead apron), wrists, genital of the operator (inside lead apron), and 6 of photoluminescence glass dosimeters were placed at eyes, thyroids, chest and genital of the patient. Exposure time and DAP values were 11.7 ± 11.8 min and 23.2 ± 26.2 Gy cm2 for electrophysiology study; 36.5 ± 42.1 min and 822.4 ± 125.5 Gy cm2 for radiofrequency catheter ablation; 16.2 ± 9.3 min and 27.8 ± 16.5 Gy cm2 for cardiac device implantation procedure, prospectively. 4591 electrophysiology studies can be conducted within the occupational exposure limit for the eyes (150 mSv), and 658-electrophysiology studies with radiofrequency catheter ablation can be carried out within the occupational exposure limit for the hands (500 mSv). 1654 cardiac device implantation procedure can be conducted within the occupational exposure limit for the eyes (150 mSv). The amounts of the operator and patient's radiation exposure were comparatively small. So, electrophysiology study, radio frequency catheter ablation and cardiac device implantation procedures are safe when performed with modern equipment and optimized protective radiation protect equipment.

Original languageEnglish
Pages (from-to)787-795
Number of pages9
JournalRadiation Effects and Defects in Solids
Volume170
Issue number10
DOIs
StatePublished - 3 Oct 2015
Externally publishedYes

Keywords

  • catheter ablation
  • electrophysiology study
  • photoluminescence glass dosimeters
  • radiation exposure

Fingerprint

Dive into the research topics of 'Radiation exposure to operator and patients during cardiac electrophysiology study, radiofrequency catheter ablation and cardiac device implantation procedures'. Together they form a unique fingerprint.

Cite this