Abstract
Severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2) and influenza viruses may pose enormous challenges to our healthcare system. We evaluated the performance of the PowerChek SARS-CoV-2, Influenza A & B Multiplex Real-time PCR Kit (PowerChek; Kogene Biotech, Seoul, Korea) in comparison with the BioFire Respiratory Panels 2 and 2.1 (RP2 and RP2.1; bioMérieux, Marcy l’Étoile, France), using 147 nasopharyngeal swabs. The limit of detection (LOD) of the PowerChek assay was determined using SARSCoV-2, influenza A, and B RNA standards. The LOD values of the PowerChek assay for SARS-CoV-2 and influenza A and B were 1.12, 1.24, and 0.61 copies/μL, respectively. The positive and negative percent agreements of the PowerChek assay compared with RP2 and RP2.1 were 97.5% (39/40) and 100% (107/107) for SARS-CoV-2; 100% (39/39) and 100% (108/108) for influenza A; and 100% (35/35) and 100% (112/112) for influenza B, respectively. The performance of the PowerChek assay was comparable to that of RP2 and RP2.1 for detecting SARS-CoV-2 and influenza A and B, suggesting its use in diagnosing SARS-CoV-2 and influenza infections.
| Original language | English |
|---|---|
| Pages (from-to) | 473-477 |
| Number of pages | 5 |
| Journal | Annals of Laboratory Medicine |
| Volume | 42 |
| Issue number | 4 |
| DOIs | |
| State | Published - 2022 |
Keywords
- Agreement
- Influenza
- Influenza A&B multiplex real-time PCR Kit
- Performance
- PowerChek SARS-CoV-2
- Real-time PCR
- SARSCoV-2