TY - JOUR
T1 - Intraoperative radiofrequency ablation using a loop internally cooled-perfusion electrode
T2 - In vitro and in vivo experiments
AU - Lee, Jeong Min
AU - Han, Joon Koo
AU - Eoh, Hong
AU - Kim, Se Hyung
AU - Lee, Jae Young
AU - Lee, Min Woo
AU - Choi, Byung Ihn
PY - 2006/4
Y1 - 2006/4
N2 - Objective. We sought to validate the efficiency of intraoperative radiofrequency ablation (RFA) using a prototype loop internally cooled-perfusion (LICP) electrode to induce coagulation in the subcapsular portion of the liver. Materials and methods. In in vitro experiments, 30 ablation regions were created using a 200-W generator and a LICP electrode featured simultaneous intraelectrode cooling and continuous flow of hypertonic saline along the shaft in explanted bovine liver. In the in vivo experiments, 26 ablation zones were created according to one of the five protocols in 10 dogs: group A, RFA using a cooled-tip electrode (n = 6); group B, RFA using a LICP electrode with 2 cm loop tip (n = 6); group C, RFA using a LICP electrode with a 3-cm loop tip (n = 6); group D, RFA using a cooled-tip electrode and Pringle maneuver (n = 4); and group E, RFA using a LICP electrode with 2 cm loop tip and Pringle maneuver (n = 4). The dimensions of the coagulation parameters were compared between the groups. Results. In the in vitro experiments, RFA using a 2- or 3-cm diameter LICP electrode (3.6 ± 0.3 cm, 3.4 ± 0.5 cm, respectively) created deeper dimensions of coagulation than did a 4-cm electrode (2.3 ± 0.2 cm; P < 0.05). In the in vivo experiments, the RFA using the LICP electrodes in porcine liver with normal perfusion induced wider but superficial ablation regions when compared to standard RFA using an internally cooled electrode (P < 0.05). However, using a Pringle maneuver, RFA with a LICP electrode created a larger volume of ablation area when compared to RFA using an internally cooled electrode with a similar range of axial diameter along the electrode axis: 30.0 ± 6.1 cm3 (group D) versus 68.5 ± 14.0 cm3 (group E; P < 0.05) Conclusions. Intraoperative RFA using the LICP electrode induced a well-defined semicircular coagulation with a 3.5-cm axial diameter in the subcapsular region of the liver. This device appears to be promising for the treatment of superficial tumors during intraoperative RFA.
AB - Objective. We sought to validate the efficiency of intraoperative radiofrequency ablation (RFA) using a prototype loop internally cooled-perfusion (LICP) electrode to induce coagulation in the subcapsular portion of the liver. Materials and methods. In in vitro experiments, 30 ablation regions were created using a 200-W generator and a LICP electrode featured simultaneous intraelectrode cooling and continuous flow of hypertonic saline along the shaft in explanted bovine liver. In the in vivo experiments, 26 ablation zones were created according to one of the five protocols in 10 dogs: group A, RFA using a cooled-tip electrode (n = 6); group B, RFA using a LICP electrode with 2 cm loop tip (n = 6); group C, RFA using a LICP electrode with a 3-cm loop tip (n = 6); group D, RFA using a cooled-tip electrode and Pringle maneuver (n = 4); and group E, RFA using a LICP electrode with 2 cm loop tip and Pringle maneuver (n = 4). The dimensions of the coagulation parameters were compared between the groups. Results. In the in vitro experiments, RFA using a 2- or 3-cm diameter LICP electrode (3.6 ± 0.3 cm, 3.4 ± 0.5 cm, respectively) created deeper dimensions of coagulation than did a 4-cm electrode (2.3 ± 0.2 cm; P < 0.05). In the in vivo experiments, the RFA using the LICP electrodes in porcine liver with normal perfusion induced wider but superficial ablation regions when compared to standard RFA using an internally cooled electrode (P < 0.05). However, using a Pringle maneuver, RFA with a LICP electrode created a larger volume of ablation area when compared to RFA using an internally cooled electrode with a similar range of axial diameter along the electrode axis: 30.0 ± 6.1 cm3 (group D) versus 68.5 ± 14.0 cm3 (group E; P < 0.05) Conclusions. Intraoperative RFA using the LICP electrode induced a well-defined semicircular coagulation with a 3.5-cm axial diameter in the subcapsular region of the liver. This device appears to be promising for the treatment of superficial tumors during intraoperative RFA.
KW - Experimental study
KW - Interventional procedures
KW - Liver
KW - Radiofrequency ablation
UR - https://www.scopus.com/pages/publications/33644913385
U2 - 10.1016/j.jss.2005.11.584
DO - 10.1016/j.jss.2005.11.584
M3 - Article
C2 - 16427086
AN - SCOPUS:33644913385
SN - 0022-4804
VL - 131
SP - 215
EP - 224
JO - Journal of Surgical Research
JF - Journal of Surgical Research
IS - 2
ER -