Fractional exhaled nitric oxide: Comparison between portable devices and correlation with sputum eosinophils

Research output: Contribution to journalArticlepeer-review

Abstract

This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO® and NObreath® concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE ≥ 3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO® (FeNONIOX-MINO) and NObreath® (FeNONObreath) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNONIOX-MINO was 0.733 (P < 0.001), and 0.751 between ISE and FeNONObreath (P < 0.001). The ROC curve found that the FeNONIOXMINO of 37.5 ppb and the FeNONObreath of 36.5 ppb identified ISE = 3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO ® and NObreath® agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE ≥ 3%.

Original languageEnglish
Pages (from-to)404-408
Number of pages5
JournalAllergy, Asthma and Immunology Research
Volume7
Issue number4
DOIs
StatePublished - 2015

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Asthma
  • Electrochemical technique
  • Eosinophils
  • Nitric oxide
  • Sputum

Fingerprint

Dive into the research topics of 'Fractional exhaled nitric oxide: Comparison between portable devices and correlation with sputum eosinophils'. Together they form a unique fingerprint.

Cite this