Abstract

Background: Recently, the iNtRON VRE vanA/vanB real-time PCR (iNtRON; iNtRON Biotechnology, Korea) assay, a multiplex real-time PCR method, was introduced. In this prospective study, we compared the iNtRON assay with the Seeplex VRE ACE detection kit (Seeplex; Seegene, Korea), a conventional multiplex PCR assay.

Methods: A chromogenic agar-based culture, in which pre-selected vancomycin-resistant enterococci (VRE) was grown and subsequently plated on blood agar with vancomycin disks, was regarded as the reference method. A total of 304 consecutive rectal swab specimens were tested for VRE by culture and by iNtRON and Seeplex PCR assays. For the PCR assays, specimens were enriched for 16-24 hr before PCR.

Results: VRE were isolated from 44 (14.5%) specimens by chromogenic agar-based culture. The clinical sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the iNtRON assay were 100% (95% confidence interval: 89.8%-100%), 99.2% (96.9%-99.9%), 95.6% (83.6%-99.2%), and 100% (98.2%-100%), respectively, while those of the Seeplex assay were 97.7% (86.2%-99.9%), 99.6% (97.5%-99.9%), 97.7% (86.2%-99.9%), and 99.6% (97.5%-99.9%), respectively. The iNtRON assay had a detection limit of 3,159 copies/μL and 13,702 copies/μL for the vanA and vanB genes, respectively. No cross-reactivity was observed in 11 non-VRE bacterial culture isolates.

Conclusions: The overall performance of the iNtRON assay was comparable to that of a chromogenic agar-based culture method for prompt identification of VRE-colonized patients in hospitals. This assay could be an alternative or supportive method for the effective control of nosocomial VRE infection.

Original languageEnglish
Pages (from-to)76-81
Number of pages6
JournalAnnals of Laboratory Medicine
Volume35
Issue number1
DOIs
StatePublished - 2015

Keywords

  • Performance
  • Real-time PCR
  • VanA
  • VanB
  • Vancomycin-resistant enterococci

Fingerprint

Dive into the research topics of 'Evaluation of the iNtRON VRE vanA/vanB real-time PCR assay for detection of vancomycin-resistant enterococci'. Together they form a unique fingerprint.

Cite this