TY - JOUR
T1 - Evaluation of maximum-likelihood position estimation with Poisson and Gaussian noise models in a small gamma camera
AU - Chung, Yong Hyun
AU - Choi, Yong
AU - Song, Tae Yong
AU - Jung, Jin Ho
AU - Cho, Gyuseong
AU - Choe, Yearn Seong
AU - Lee, Kyung Han
AU - Kim, Sang Eun
AU - Kim, Byung Tae
PY - 2004/2
Y1 - 2004/2
N2 - It has been reported that maximum-likelihood position estimation (MLPE) algorithms offer advantages of improved spatial resolution and linearity over conventional Anger algorithm in gamma cameras. While the fluctuation of photon measurements is more accurately described by Poisson than Gaussian distribution model, the likelihood function of a scintillation event assumed to be Gaussian could be more easily implemented and might provide more consistent outcomes than Poisson-based MLPE. The purpose of this study is to evaluate the performances of the noise models, Poisson and Gaussian, in MLPE for the localization of photons in a small gamma camera (SGC) using NaI(Tl) plate and PSPMT. The SGC consisted of a single NaI(Tl) crystal, 10 cm in diameter and 6 mm thick, optically coupled to a PSPMT (Hamamatsu R3292-07). The PSPMT was read out using a resistive charge divider, which multiplexes 28(X) by 28(Y) cross wire anodes into four channels. Poisson and Gaussian-based MLPE methods have been implemented using experimentally measured detector response functions (DRF). The averaged intrinsic spatial resolutions were 3.14, 3.09, and 2.88 mm, the integral uniformities were 15.3%, 12.3%, and 11.4%, and the averaged linearities were 0.75, 0.33, and 0.22 mm for Anger logic, Poisson, and Gaussian-based MLPE, respectively. MLPEs considerably improved linearity and uniformity compared to Anger logic. Gaussian-based MLPE, which is easy to implement, allowed to obtain better linearity and uniformity performances than the Poisson-based MLPE.
AB - It has been reported that maximum-likelihood position estimation (MLPE) algorithms offer advantages of improved spatial resolution and linearity over conventional Anger algorithm in gamma cameras. While the fluctuation of photon measurements is more accurately described by Poisson than Gaussian distribution model, the likelihood function of a scintillation event assumed to be Gaussian could be more easily implemented and might provide more consistent outcomes than Poisson-based MLPE. The purpose of this study is to evaluate the performances of the noise models, Poisson and Gaussian, in MLPE for the localization of photons in a small gamma camera (SGC) using NaI(Tl) plate and PSPMT. The SGC consisted of a single NaI(Tl) crystal, 10 cm in diameter and 6 mm thick, optically coupled to a PSPMT (Hamamatsu R3292-07). The PSPMT was read out using a resistive charge divider, which multiplexes 28(X) by 28(Y) cross wire anodes into four channels. Poisson and Gaussian-based MLPE methods have been implemented using experimentally measured detector response functions (DRF). The averaged intrinsic spatial resolutions were 3.14, 3.09, and 2.88 mm, the integral uniformities were 15.3%, 12.3%, and 11.4%, and the averaged linearities were 0.75, 0.33, and 0.22 mm for Anger logic, Poisson, and Gaussian-based MLPE, respectively. MLPEs considerably improved linearity and uniformity compared to Anger logic. Gaussian-based MLPE, which is easy to implement, allowed to obtain better linearity and uniformity performances than the Poisson-based MLPE.
KW - Position measurement
KW - Position sensitive PMT
KW - Small gamma camera
UR - https://www.scopus.com/pages/publications/2342637068
U2 - 10.1109/TNS.2003.823053
DO - 10.1109/TNS.2003.823053
M3 - Article
AN - SCOPUS:2342637068
SN - 0018-9499
VL - 51
SP - 101
EP - 104
JO - IEEE Transactions on Nuclear Science
JF - IEEE Transactions on Nuclear Science
IS - 1 I
ER -