T1 - Estimation of photoneutron yield in linear accelerator with different collimation systems by Geant4 and MCNPX simulation codes
AU - Kim, Yoon Sang
AU - Khazaei, Zeinab
AU - Ko, Junho
AU - Afarideh, Hossein
AU - Ghergherehchi, Mitra
N1 - Publisher Copyright:
© 2016 Institute of Physics and Engineering in Medicine.
PY - 2016/3/15
Y1 - 2016/3/15
N2 - At present, the bremsstrahlung photon beams produced by linear accelerators are the most commonly employed method of radiotherapy for tumor treatments. A photoneutron source based on three different energies (6, 10 and 15 MeV) of a linac electron beam was designed by means of Geant4 and Monte Carlo N-Particle eXtended (MCNPX) simulation codes. To obtain maximum neutron yield, two arrangements for the photo neutron convertor were studied: (a) without a collimator, and (b) placement of the convertor after the collimator. The maximum photon intensities in tungsten were 0.73, 1.24 and 2.07 photon/e at 6, 10 and 15 MeV, respectively. There was no considerable increase in the photon fluence spectra from 6 to 15 MeV at the optimum thickness between 0.8 mm and 2 mm of tungsten. The optimum dimensions of the collimator were determined to be a length of 140 mm with an aperture of 5 mm X 70 mm for iron in a slit shape. According to the neutron yield, the best thickness obtained for the studied materials was 30 mm. The number of neutrons generated in BeO achieved the maximum value at 6 MeV, unlike that in Be, where the highest number of neutrons was observed at 15 MeV. Statistical uncertainty in all simulations was less than 0.3% and 0.05% for MCNPX and the standard electromagnetic (EM) physics packages of Geant4, respectively. Differences among spectra in various regions are due to various cross-section and stopping power data and different simulations of the physics processes.
AB - At present, the bremsstrahlung photon beams produced by linear accelerators are the most commonly employed method of radiotherapy for tumor treatments. A photoneutron source based on three different energies (6, 10 and 15 MeV) of a linac electron beam was designed by means of Geant4 and Monte Carlo N-Particle eXtended (MCNPX) simulation codes. To obtain maximum neutron yield, two arrangements for the photo neutron convertor were studied: (a) without a collimator, and (b) placement of the convertor after the collimator. The maximum photon intensities in tungsten were 0.73, 1.24 and 2.07 photon/e at 6, 10 and 15 MeV, respectively. There was no considerable increase in the photon fluence spectra from 6 to 15 MeV at the optimum thickness between 0.8 mm and 2 mm of tungsten. The optimum dimensions of the collimator were determined to be a length of 140 mm with an aperture of 5 mm X 70 mm for iron in a slit shape. According to the neutron yield, the best thickness obtained for the studied materials was 30 mm. The number of neutrons generated in BeO achieved the maximum value at 6 MeV, unlike that in Be, where the highest number of neutrons was observed at 15 MeV. Statistical uncertainty in all simulations was less than 0.3% and 0.05% for MCNPX and the standard electromagnetic (EM) physics packages of Geant4, respectively. Differences among spectra in various regions are due to various cross-section and stopping power data and different simulations of the physics processes.
KW - collimator
KW - linear accelerator
KW - Monte Carlo simulation code
KW - neutron yield
KW - photoneutron convertor
UR - https://www.scopus.com/pages/publications/84962325642
U2 - 10.1088/0031-9155/61/7/2762
DO - 10.1088/0031-9155/61/7/2762
M3 - Article
C2 - 26975304
AN - SCOPUS:84962325642
SN - 0031-9155
VL - 61
SP - 2762
EP - 2779
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 7