Efficacy and Safety of AMPLATZER Vascular Plug Type IV for Embolization of Pulmonary Arteriovenous Malformations

Sang Yub Lee, Jongmin Lee, Young Hwan Kim, Ung Rae Kang, Jung Guen Cha, Jaehee Lee, Seung Ick Cha, Chang Ho Kim

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Purpose: To assess the efficacy and safety of the AMPLATZER Vascular Plug type IV for pulmonary arteriovenous malformation (PAVM) treatment. Materials and Methods: Between June 2013 and January 2018, 13 patients with 26 PAVMs were treated with the type IV AVP. Patients without follow-up computed tomography (CT) were excluded. Technical success was defined as flow occlusion on angiography. Plug-to-sac distance was measured on angiographic images. Feeding artery and venous sac diameter changes were measured on preprocedural and follow-up CT. Successful embolization was defined as > 70% sac size regression. Procedure time, device migration, and complications were evaluated. Results: Nine female patients (mean age, 49 y; range, 40–71 y) with 19 PAVMs were enrolled. Four patients with 7 PAVMs were lost to follow-up. Nineteen PAVMs were treated in 11 sessions, and the mean procedure time was 29 min. The technical success rate was 100%. Mean feeding artery diameter was 3.1 mm ± 0.7 (range, 2.1–4.9 mm). Mean plug-to-sac distance was 5.4 mm ± 4.9 (range, 0–13.3 mm). The mean CT follow-up period was 14 months ± 7 (range, 6–30 mo). Sixteen of 19 PAVMs (84%) were successfully embolized. Minor complications (tachycardia and chest discomfort) arose in 2 of 11 sessions. No device migrations or major complications occurred. Conclusions: The type IV AVP showed an 84% treatment success rate based on 70% sac size regression criteria in small PAVMs. There were no device migrations or major complications.

Original languageEnglish
Pages (from-to)1082-1088
Number of pages7
JournalJournal of Vascular and Interventional Radiology
Volume30
Issue number7
DOIs
StatePublished - Jul 2019
Externally publishedYes

Fingerprint

Dive into the research topics of 'Efficacy and Safety of AMPLATZER Vascular Plug Type IV for Embolization of Pulmonary Arteriovenous Malformations'. Together they form a unique fingerprint.

Cite this