Abstract
We evaluated the effectiveness of the predictive low-glucose suspend (PLGS) algorithm in the DIA:CONN G8. Forty people with type 1 diabetes mellitus (T1DM) who used a DIA:CONN G8 for at least 2 months with prior experience using pumps without and with PLGS were retrospectively analyzed. The objective was to assess the changes in time spent in hypoglycemia (percent of time below range [%TBR]) before and after using PLGS. The mean age, sensor glucose levels, glucose threshold for suspension, and suspension time were 31.1±22.8 years, 159.7±23.2 mg/dL, 81.1±9.1 mg/dL, and 111.9±79.8 min/day, respectively. Overnight %TBR <70 mg/dL was significantly reduced after using the algorithm (differences=0.3%, from 1.4%±1.5% to 1.1%±1.2%, P=0.045). The glycemia risk index (GRI) improved significantly by 4.2 (from 38.8±20.9 to 34.6±19.0, P=0.002). Using the PLGS did not result in a change in the hyperglycemia metric (all P>0.05). Our findings support the PLGS in DIA:CONN G8 as an effective algorithm to improve night-time hypoglycemia and GRI in people with T1DM.
| Original language | English |
|---|---|
| Pages (from-to) | 144-149 |
| Number of pages | 6 |
| Journal | Diabetes and Metabolism Journal |
| Volume | 49 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 2025 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Blood glucose
- Continuous glucose monitoring
- Diabetes mellitus, type 1
- Hypoglycemia
- Insulin infusion systems
Fingerprint
Dive into the research topics of 'Effectiveness of Predicted Low-Glucose Suspend Pump Technology in the Prevention of Hypoglycemia in People with Type 1 Diabetes Mellitus: Real-World Data Using DIA:CONN G8'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver