Skip to main navigation Skip to search Skip to main content

Effect of the working and counter/quasi-reference electrode relative area ratio of silver sensor electrodes on voltammetric detection of Pb(II)

  • Sungkyunkwan University
  • Kangwon National University

Research output: Contribution to journalArticlepeer-review

Abstract

The electrochemical properties of the voltammetric Pb(II) sensor were investigated by considering the influence of the relative geometric area ratio (working electrode: counter/quasi-reference electrode = 3.20 (R3), 10.45 (R10), and 22.13 (R22)) to determine the trace levels of Pb(II) in solution. The limits of detection value were 6.67 nM, 8.61 nM, and 10.6 nM for R3, R10, and R22, respectively and were fully satisfied according to the guideline value of the World Health Organization for Pb(II) in drinking water. In addition, more than 90% of the initial activity for the reduction peak current remained up to 8 times for R3, 7 times for R10, and 5 times for R22. These results demonstrated that the geometric area ratio of the working electrode with the integrated silver counter to the quasi-reference electrode was significant to suggest that the geometric area ratio is an influencing factor in the electrochemical sensor performance.

Original languageEnglish
Pages (from-to)67-70
Number of pages4
JournalJournal of Industrial and Engineering Chemistry
Volume81
DOIs
StatePublished - 25 Jan 2020

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Acid mine drainage
  • Geometric area ratio
  • Lead
  • Pb(II)
  • Silver electrode
  • Voltammetric sensor

Fingerprint

Dive into the research topics of 'Effect of the working and counter/quasi-reference electrode relative area ratio of silver sensor electrodes on voltammetric detection of Pb(II)'. Together they form a unique fingerprint.

Cite this