Abstract

Purpose: In stereotactic radiotherapy using X-Knife system, the commercially supplied collimator cone system had a few mechanical limitations. The authors have developed new collimator cones to overcome these limitations and named them 'SMC type' collimator cones. Methods: We made use of cadmium-free cerrobend alloy within the stainless steel cylinder housing. We made nine cones of relatively larger sizes (3.0 cm to 7.0 cm in diameter) and of shorter length with bigger clearance from the isocenter than the commercial cones. The cone housing and the collimator cones were designed to insert into the wedge mount of the gantry head to enable double-exposure linac-gram taking. Results: The mechanical accuracy of pointing to the isocenter was tested by ball test and cone rotation test, and the dosimetric measurements were performed, all of which were with satisfactory results. A new innovative quality assurance procedure using linac-grams on the patients at the actual treatment setup was attempted after taking 10 sets of AP and lateral linac-grams and the overall mechanical isocenter accuracy was excellent (average error = 0.4 ± 0.2 mm). Conclusions: We have developed the SMC type collimator cone system mainly for fractionated stereotactive radiation therapy use with our innovative ideas. The new cones' mechanical accuracy and physical properties were satisfactory for clinical use, and the verification of the isocenter accuracy on the actual treatment setup has become possible.

Original languageEnglish
Pages (from-to)435-438
Number of pages4
JournalInternational Journal of Radiation Oncology Biology Physics
Volume44
Issue number2
DOIs
StatePublished - May 1999

Keywords

  • Collimator cones
  • Fractionated stereotactic radiation therapy
  • Quality assurance