Compressed sensing for fMRI: Feasibility study on the acceleration of non-EPI fMRI at 9.4T

Paul Kyu Han, Sung Hong Park, Seong Gi Kim, Jong Chul Ye

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

Conventional functional magnetic resonance imaging (fMRI) technique known as gradient-recalled echo (GRE) echo-planar imaging (EPI) is sensitive to image distortion and degradation caused by local magnetic field inhomogeneity at high magnetic fields. Non-EPI sequences such as spoiled gradient echo and balanced steady-state free precession (bSSFP) have been proposed as an alternative high-resolution fMRI technique; however, the temporal resolution of these sequences is lower than the typically used GRE-EPI fMRI. One potential approach to improve the temporal resolution is to use compressed sensing (CS). In this study, we tested the feasibility of k-t FOCUSS - one of the high performance CS algorithms for dynamic MRI - for non-EPI fMRI at 9.4T using the model of rat somatosensory stimulation. To optimize the performance of CS reconstruction, different sampling patterns and k-t FOCUSS variations were investigated. Experimental results show that an optimized k-t FOCUSS algorithm with acceleration by a factor of 4 works well for non-EPI fMRI at high field under various statistical criteria, which confirms that a combination of CS and a non-EPI sequence may be a good solution for high-resolution fMRI at high fields.

Original languageEnglish
Article number131926
JournalBioMed Research International
Volume2015
DOIs
StatePublished - 2015

Fingerprint

Dive into the research topics of 'Compressed sensing for fMRI: Feasibility study on the acceleration of non-EPI fMRI at 9.4T'. Together they form a unique fingerprint.

Cite this