T1 - Comparing feasibility of low-tube-voltage protocol with low-iodine-concentration contrast and high-tube-voltage protocol with high-iodine-concentration contrast in coronary computed tomography angiography
AU - Cha, Min Jae
AU - Kim, Sung Mok
AU - Ahn, Tae Ran
AU - Choe, Yeon Hyeon
N1 - Publisher Copyright:
© 2020 Cha et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2020/7
Y1 - 2020/7
N2 - Background To investigate the feasibility of a low tube voltage (80 kVp) protocol with low concentration contrast media (CM) (iodixanol 320 mgl/ml) as compared with a high tube voltage (100 kVp) protocol with high concentration CM (iomeprol 400 mgl/ml) in coronary CT angiography (CCTA) for patients with body mass index less than 30. Materials and methods A total of 93 patients were randomly assigned into three groups and underwent CCTA as follows: Group A) 100 kVp, 100–350 mAs, 400 mgl/ml CM at 4ml/s, and reconstructed with filtered back projection; Group B and C) 80 kVp, 100–450 mAs, 320 mgl/ml CM at 4 ml/s and 5 ml/s, respectively and reconstructed with iterative reconstruction. Objective and subjective image quality (IQ) was analyzed. Results The image noise, intravascular attenuation, signal-to-noise ratio and contrast-to-noise ratio of major coronary arteries did not differ significantly among three groups. Subjective IQ analyses on vascular attenuation and image noise did not differ significantly, either (all of p > 0.05). Qualitative IQ of Group B and C was non-inferior to that of Group A. Substantial reduction of radiation exposure was achieved in group B (2.60 ± 0.48 mSv) and C (2.72 ± 0.54 mSv), compared with group A (3.58 ± 0.67 mSv) (p < 0.05). Conclusion CCTA at 80 kVp with 320 mgl/ml CM and iterative reconstruction is feasible, achieving radiation dose reduction, while preserving IQ.
AB - Background To investigate the feasibility of a low tube voltage (80 kVp) protocol with low concentration contrast media (CM) (iodixanol 320 mgl/ml) as compared with a high tube voltage (100 kVp) protocol with high concentration CM (iomeprol 400 mgl/ml) in coronary CT angiography (CCTA) for patients with body mass index less than 30. Materials and methods A total of 93 patients were randomly assigned into three groups and underwent CCTA as follows: Group A) 100 kVp, 100–350 mAs, 400 mgl/ml CM at 4ml/s, and reconstructed with filtered back projection; Group B and C) 80 kVp, 100–450 mAs, 320 mgl/ml CM at 4 ml/s and 5 ml/s, respectively and reconstructed with iterative reconstruction. Objective and subjective image quality (IQ) was analyzed. Results The image noise, intravascular attenuation, signal-to-noise ratio and contrast-to-noise ratio of major coronary arteries did not differ significantly among three groups. Subjective IQ analyses on vascular attenuation and image noise did not differ significantly, either (all of p > 0.05). Qualitative IQ of Group B and C was non-inferior to that of Group A. Substantial reduction of radiation exposure was achieved in group B (2.60 ± 0.48 mSv) and C (2.72 ± 0.54 mSv), compared with group A (3.58 ± 0.67 mSv) (p < 0.05). Conclusion CCTA at 80 kVp with 320 mgl/ml CM and iterative reconstruction is feasible, achieving radiation dose reduction, while preserving IQ.
UR - https://www.scopus.com/pages/publications/85088156744
U2 - 10.1371/journal.pone.0236108
DO - 10.1371/journal.pone.0236108
M3 - Article
C2 - 32673356
AN - SCOPUS:85088156744
SN - 1932-6203
VL - 15
JO - PLoS ONE
JF - PLoS ONE
IS - 7 July
M1 - e0236108
ER -