Skip to main navigation Skip to search Skip to main content

Clinical Evaluation of a Hydrophobic Intraocular Lens Using a Preloaded Automated Injector in a Korean Population

  • Hong Kyun Kim
  • , Kyoung Yul Seo
  • , Kyung Chul Yoon
  • , Chul Young Choi
  • , Tae Young Chung
  • , Joon Young Hyon
  • , Alexis Rendon
  • , Hyun Seung Kim
  • Kyungpook National University
  • Yonsei University
  • Chonnam National University
  • Kangbuk Samsung Hospital
  • Seoul National University
  • Alcon Vision LLC
  • The Catholic University of Korea

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: This study assessed post-market clinical outcomes of the Clareon monofocal intraocular lens (IOL) preloaded in the AutonoMe Delivery System in a real-world setting of Korean patients. Methods: This prospective, multicenter, single-arm study in Korea was conducted from July 2020 to December 2021. Patients were ≥20 years old with unilateral or bilateral cataracts who received Clareon IOLs (CNA0T0) preloaded in an automated injector system. Best corrected distance visual acuity (BCDVA) and uncorrected distance visual acuity (UCDVA) were evaluated under photopic conditions. Surgeon delivery system preference was assessed using a survey questionnaire. Glistenings, surface haze, adverse events, posterior capsule opacification (PCO), and Nd:YAG capsulotomy rates were also assessed during the 12-month postoperative follow-up. Results: Mean ± SD monocular BCDVA was 0.02 ± 0.11 and 0.00 ± 0.10 logMAR at 1 month and 12 months, respectively. BCDVA of 0.2 logMAR or better was achieved by 94.4% and 99.1% of eyes at 1 month and 12 months after implantation, respectively. Mean monocular UCDVA was 0.11 ± 0.14 and 0.07 ± 0.13 logMAR at 1 month and 12 months, respectively. UCDVA of 0.3 logMAR or better was achieved by 97.4% of eyes at 12 months after implantation. Preparation of the automated injector system was rated as “very easy” or “easy” and CNA0T0 IOL delivery was rated as “very controllable” or “controllable” by all surgeons. Only grade 0 glistenings and no surface haze were observed during the 12-month follow-up. No clinically significant PCO or Nd:YAG capsulotomy were reported throughout the study; clinically nonsignificant PCO was reported in 23% of eyes. Conclusion: This 12-month real-world study of the CNA0T0 IOL and the automated injector system demonstrated excellent visual outcomes and high surgeon satisfaction.

Original languageEnglish
Pages (from-to)3353-3363
Number of pages11
JournalClinical Ophthalmology
Volume17
DOIs
StatePublished - 2023
Externally publishedYes

Keywords

  • glistenings
  • intraocular lens delivery
  • surface haze
  • surgeon’s preference questionnaire
  • visual acuity

Fingerprint

Dive into the research topics of 'Clinical Evaluation of a Hydrophobic Intraocular Lens Using a Preloaded Automated Injector in a Korean Population'. Together they form a unique fingerprint.

Cite this