Abstract
Background: We previously reported unexpected early failures of total hip arthroplasty (THA) utilizing a 28-mm Ultima metal-on-metal (MoM) articulation. However, long-term results of small-diameter MoM THAs still remain unclear. The purpose of this study was to evaluate the clinical and radiographic outcomes of the previously reported cohort at a minimum follow-up of 15 years. Methods: The original cohort consisted of 171 primary THAs (167 patients) using a 28-mm MoM articulation performed between April 2000 and March 2002. Of these, 130 hips (126 patients) were reviewed at an average follow-up of 17.1 (range, 15-18) years. Clinical score, complications, presence of osteolytic lesion, serum metal ion concentrations, and implant survivorships were evaluated. Results: The mean Harris Hip Score improved from 44.9 points preoperatively to 90.5 points at the latest follow-up. During the entire period since the original surgery, a total of 5 revisions (3.8%) were associated with adverse reaction to metal debris (ARMD). The last revision surgery for symptomatic ARMD was performed at 6 years postoperatively. The implant survivorships with an end point of revision for aseptic failure and for any reason were 95.4% and 93.8%, respectively. Radiographic osteolytic lesions were detected in 27 hips (20.8%). The average serum metal concentration was 2.50 (range, 0.12-9.86) μg/L for cobalt and 2.81 (range, 0.82-12.3) μg/L for chromium. Conclusions: THA using a 28-mm MoM articulation showed favorable long-term outcomes with a relatively high revision-free survival rate. There was no significant symptomatic ARMD after the last acetabular revision performed at 6 years postoperatively.
| Original language | English |
|---|---|
| Pages (from-to) | 1387-1394 |
| Number of pages | 8 |
| Journal | Journal of Arthroplasty |
| Volume | 34 |
| Issue number | 7 |
| DOIs | |
| State | Published - Jul 2019 |
Keywords
- 28-mm
- fifteen-year results
- metal-on-metal
- total hip arthroplasty
- Ultima
Fingerprint
Dive into the research topics of 'A Minimum of 15-Year Results of Cementless Total Hip Arthroplasty Using a 28-mm Metal-On-Metal Articulation'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver