A high-precision angular control system for HIFU calibration

  • Donghee Park
  • , Jingam Park
  • , Hansung Kim
  • , Chi Hyun Kim
  • , Tae Young Han
  • , Hyunjin Park
  • , Jongbum Seo

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

A design of high-precision angular position control system for calibrating high intensity focused ultrasound (HIFU) is presented with alignment procedures. Two independent angular controls are achieved by combining a worm gear and a belt gear system. The proposed system verifies alignment by comparing simulation data and experimental data with three different transducers and two different types of hydrophones. The performance of the proposed system is compared to that of a commercial system. The results indicate that the proposed system provides high precision angular alignment (e.g., <0.01 radians) with robust reproducibility regardless of the hydrophone type.

Original languageEnglish
Pages (from-to)45-52
Number of pages8
JournalUltrasonics
Volume53
Issue number1
DOIs
StatePublished - Jan 2013

Keywords

  • Acoustic field
  • High intensity focused ultrasound
  • High precision alignment
  • Ultrasound field measurement

Fingerprint