6 MV X-band linear accelerator for stereotactic body radiation therapy

  • Donghyup Ha
  • , Seung Hyun Lee
  • , Mitra Ghergherehchi
  • , Hyojeong Choi
  • , Ho Namgoong
  • , Jongchul Lee
  • , Jong Seo Chai

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

This paper presents the development process of a high-dose X-band linear accelerator (LINAC) for radiotherapy machines, specifically focusing on stereotactic body radiation therapy (SBRT). The development process includes a multi-physics design, approach for the radio frequency cavity in the LINAC, with an emphasis on minimizing beam size and maximizing shunt impedance of the LINAC. Commissioning tests were conducted to evaluate the performance of the LINAC, which involved measurements of beam energy, beam size, beam current, and X-ray dose rate.

Original languageEnglish
Pages (from-to)4502-4511
Number of pages10
JournalNuclear Engineering and Technology
Volume56
Issue number11
DOIs
StatePublished - Nov 2024
Externally publishedYes

Keywords

  • LINAC
  • Radiation therapy
  • Stereotactic body radiation therapy
  • X-band

Fingerprint

Dive into the research topics of '6 MV X-band linear accelerator for stereotactic body radiation therapy'. Together they form a unique fingerprint.

Cite this