TY - JOUR
T1 - 2-Dimensional shear wave elastography
T2 - Interobserver agreement and factors related to interobserver discrepancy
AU - Yoon, Kibo
AU - Jeong, Woo Kyoung
AU - Kim, Yongsoo
AU - Kim, Min Yeong
AU - Kim, Tae Yeob
AU - Sohn, Joo Hyun
N1 - Publisher Copyright:
© 2017 Yoon et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2017/4
Y1 - 2017/4
N2 - Purpose: To evaluate the interobserver reproducibility of two-dimensional shear wave elastography (2D-SWE) in measuring liver stiffness (LS) and to investigate factors related to liver 2D-SWE. Materials and methods: A prospective study was conducted between August 2011 and August 2012 in rheumatoid arthritis patients who had been treated with methotrexate. Interobserver reproducibility of 2D-SWE was evaluated, and the relationship between interobserver difference in LS and related factors was analyzed using linear regression analyses. We considered age, sex, alanine transaminase, cholesterol, body mass index (BMI), and waist circumference as clinical factors, and the mean value of standard deviation (SDM), its difference between two examiners, mean diameter of the regions of interest (ROIM), and its difference in the elasticity map as investigation factors. The cut-off value for significant factors to predict interobserver discrepancies in LS-based fibrosis stage was also inspected. Results: In total, 176 patients were enrolled. The intraclass correlation coefficient between the two examiners was 0.784. In the univariate analysis, SDM and ROIM were independently associated with interobserver differences in LS as well as BMI, waist circumference, and the difference of ROI, but SDM and ROIM were the only ones significantly related in multivariate analysis (p<0.001 and p = 0.021, respectively). The best cut-off value for SDM in predicting interobserver discrepancy in LS-based fibrosis stage was 1.4. Conclusions: Interobserver reproducibility of 2D-SWE for measuring LS was good and SDM was the most significantly associated factor with interobserver differences in LS and interobserver discordance in LS-based fibrosis stage.
AB - Purpose: To evaluate the interobserver reproducibility of two-dimensional shear wave elastography (2D-SWE) in measuring liver stiffness (LS) and to investigate factors related to liver 2D-SWE. Materials and methods: A prospective study was conducted between August 2011 and August 2012 in rheumatoid arthritis patients who had been treated with methotrexate. Interobserver reproducibility of 2D-SWE was evaluated, and the relationship between interobserver difference in LS and related factors was analyzed using linear regression analyses. We considered age, sex, alanine transaminase, cholesterol, body mass index (BMI), and waist circumference as clinical factors, and the mean value of standard deviation (SDM), its difference between two examiners, mean diameter of the regions of interest (ROIM), and its difference in the elasticity map as investigation factors. The cut-off value for significant factors to predict interobserver discrepancies in LS-based fibrosis stage was also inspected. Results: In total, 176 patients were enrolled. The intraclass correlation coefficient between the two examiners was 0.784. In the univariate analysis, SDM and ROIM were independently associated with interobserver differences in LS as well as BMI, waist circumference, and the difference of ROI, but SDM and ROIM were the only ones significantly related in multivariate analysis (p<0.001 and p = 0.021, respectively). The best cut-off value for SDM in predicting interobserver discrepancy in LS-based fibrosis stage was 1.4. Conclusions: Interobserver reproducibility of 2D-SWE for measuring LS was good and SDM was the most significantly associated factor with interobserver differences in LS and interobserver discordance in LS-based fibrosis stage.
UR - https://www.scopus.com/pages/publications/85017570307
U2 - 10.1371/journal.pone.0175747
DO - 10.1371/journal.pone.0175747
M3 - Article
C2 - 28414822
AN - SCOPUS:85017570307
SN - 1932-6203
VL - 12
JO - PLoS ONE
JF - PLoS ONE
IS - 4
M1 - e0175747
ER -